A pharmacy technician is asked to prepare a non-sterile compound that requires the use of a Class A torsion balance. The formula calls for 15 mg of a potent active pharmaceutical ingredient (API). The minimum weighable quantity (MWQ) for the balance is 120 mg. Which of the following is the most appropriate action for the technician?

  1. AWeigh the 15 mg directly on the Class A balance, as it is a potent ingredient.
  2. BUse a digital analytical balance instead, as it is more accurate for small quantities.
  3. CRound up the required quantity to 120 mg to meet the MWQ.
  4. DPerform an aliquot weighing to obtain the required 15 mg.
Show answer & explanation

Correct answer: D. Perform an aliquot weighing to obtain the required 15 mg.

Since 15 mg is below the minimum weighable quantity (MWQ) of 120 mg for the Class A torsion balance, direct weighing would lead to an unacceptable error. The most appropriate action is to use aliquot weighing, which involves weighing a larger, diluting, and then taking a portion (aliquot) that contains the desired amount of the API.

Why the other options are wrong

  • A. Directly weighing less than the MWQ results in significant error and is unsafe.
  • B. While a digital balance might be more accurate, the question specifies using a Class A torsion balance, and aliquot method is designed for this scenario.
  • C. Rounding up to 120 mg would result in a significantly incorrect dose and is dangerous.

Aliquot Weighing

A technique used to accurately measure quantities of an ingredient that are below the minimum weighable quantity (MWQ) of a balance.

  • Involves weighing a larger amount of the ingredient mixed with an inert diluent.
  • A portion (aliquot) of this mixture is then taken to provide the desired amount of the ingredient.
  • Ensures accuracy when using balances with specific MWQs.

Memory trick: Small Amounts Need Aliquot Accuracy